| Max. Load Capacity | 10kN |
| Crosshead | Digital display (display resolution: 0.001 mm) |
| Position | |
| Detection:Display Method | |
| Power Requirements | Supply voltage fluctuations within ±10% of the set value. |
| D-class (100 Ω max.) grounding resistance. | |
| Power Requirements | Single phase 100/120/220/240 V AC |
| (switching type) 50/60 Hz 1.2 kVA | |
| Weight | 85kg |
| Dimensions | W653×D520× |
| H1603mm | |
| Accessories | Load cell (with CAL cable), Power cable (2.5 m),turning rod, cable clamps, instruction manual |
| Standard Functions | • Analog recorder (option) output |
| • Dataletty (option) output*5 | |
| • Automatic test force / stress control (Autotuning) | |
| • Automatic strain control (Autotuning) *4 | |
| • Test force auto-zeroing | |
| • Test force auto-calibration | |
| • Break detection, auto-return | |
| • Load cell overload detection | |
| • Touch-load detection function | |
| Standard Functions | • External analog input (2 channels) *4 |
| • External digital input (2 channels) *4 | |
| Standard Functions | • Automatic reading of load cell characteristic values |
| • Test force display, stress display, stroke display,position display | |
| • External analog output (2 channels) | |
| Sampling Speed | 1000Hz max. *4 |
| Crosshead | ±0.1% indicated value or ±0.01 mm, whichever is larger |
| Position | |
| Detection:Positional Accuracy | |
| Crosshead | Optical encoder |
| Position | |
| Detection:Measurement Method | |
| Lineup of Models Differing byLoad Cell Capacity | 1N /2N /5N /10N /20N /50N |
| 100N /500N | |
| 1kN/ 5kN / 10kN | |
| Effective Test Width | 425mm |
| Crosshead – Table | 1200mm |
| Distance(Tensile stroke) *3 | (760mm, MWG) |
| Crosshead Speed and | To load cell capacity across entire speed range |
| Permitted Test Force | |
| Crosshead Speed | ±0.1 |
| Accuracy*2 | |
| Crosshead : Max. Return Speed | 1500mm/min |
| Crosshead : Speed Range | 0.001 to 1000 mm/min (stepless) |
| Measurement : Calibration | Automatic test force calibration: select tensile, compression, or |
| tensile and compression | |
| Measurement : Accuracy : Standard-Precision Type (1/500, ±1%) | Within ±1% indicated test force (at 1/500 to 1/1 load cell rating) |
| Conforms to EN 10002-2 Grade 1, ISO 7500-1 Class 1, | |
| BS 1610 Class 1, and ASTM E4, JIS B7721 Class 1. *1 | |
| Force:Accuracy : High-Precision Type | Within ±0.5% indicated test force (at 1/500 to 1/1 load cell rating) |
| (1/500, ±0.5%) | Conforms to EN 10002-2 Grade 0.5, ISO 7500-1 Class 0.5, |
| BS 1610 Class 0.5, and ASTM E4, JIS B7721 Class 0.5. *1 | |
| Loading Method | Direct high-precision constant speed strain control method via non-backlash ball-screw drive |
| Operating Environment | Temperature: 5 °C to 40 °C; Humidity: 20% to 80% (no condensation) |
| Floor vibrations: frequency 10 Hz max., amplitude 5 μm max. | |





Bình luận